Micro Motion F100 Flow and Density Meter
Model:F100S129CCAZEZZZZ
Manufacturer: Micro Motion
Size: 100*80*50CM
Detail Specification: Flow and Density Meter
Comment:wooden case
Payment:T/T
HS CODE:9026201090
Delivery time: 8-10 weeks
product details
Micro Motion F100 Flow and Density Meter
Micro Motion F100S130CQBAEZZZZ Coriolis Flow and Density Meter — 1 in (DN25) | 316L Stainless Steel
Meta Title: Micro Motion F100S130CQBAEZZZZ Coriolis Flow Meter | 1 in DN25 316L | F100 Series
Meta Description: Micro Motion F100S130CQBAEZZZZ F-Series Coriolis flow and density meter. 1 in (DN25), 316L stainless steel, high-temperature option. ±0.20% mass flow accuracy. Get specs, configuration details, and application guidance.
Product Overview
The Micro Motion F100S130CQBAEZZZZ is a 1-inch (DN25) Coriolis flow and density meter from Emerson’s F-Series family. The base model F100S denotes a compact, drainable sensor constructed from 316L stainless steel — one of two wetted-material options across the F-Series, alongside C-22 nickel alloy used in the F100H, F100B, and F100P variants.
This configuration delivers direct mass flow, volume flow, density, and temperature measurement from a single device, eliminating the need for separate instruments in tight, drainable process lines. The F100 platform is purpose-built for applications where space is constrained but measurement accuracy cannot be compromised.
F100S129CCAZEZZZZ
| F100S129 | Stainless steel |
| 129 | 1 in CL300 ASME B16.5 F316/F316L Weld neck flange Raised face |
| C | Code descriptions:Compact case |
| C | Electronics interface:For integrally mounted 1700 or 2700 transmitter |
| A | Conduit connections:0.75 in (19 mm) NPT — no gland |
| Z | Approvals:ATEX – Equipment Category 2 (Zone 1) / PED compliant |
| E | Languages:English installation manual |
| Z | Additional standard approvals:No additional standard approval options selected; does not apply to F100P |
| Z | Calibration:±0.20% mass and 2 kg/m³ (0.002 g/cm³) density calibration |
| Z | Measurement application software (all models):No measurement application software |
| Z | Factory options;Standard product |
The standard mass flow accuracy is ±0.20% of rate, with optional calibration upgrades available at ±0.15% and ±0.10% for applications requiring tighter process control. Density accuracy is specified at ±0.002 g/cm³ under standard conditions, with an enhanced ±0.001 g/cm³ option. The meter’s maximum working pressure of 1,450 psig (100 barg) positions it for moderate-to-high pressure liquid and gas service.
F-Series Model Nomenclature: S, A, H, B, and P Explained
Understanding the suffix letter is critical when specifying an F100 meter for your application:
F100S — 316L stainless steel, standard temperature range. Ambient operating range: -61 °F to 149 °F (-52.0 °C to 65.0 °C). Suitable for most liquid hydrocarbon, chemical, and general process applications.
F100A — 316L stainless steel, high-temperature version. Designed for processes where elevated fluid temperatures exceed the standard S-model envelope.
F100H — Nickel alloy C22, standard temperature. C22 provides superior corrosion resistance in aggressive chemical environments where 316L would degrade.
F100B — Nickel alloy C22, high-temperature version.
F100P — Nickel alloy C22, high-pressure version. Maximum working pressure rated at 6,250 psig (431 barg) — more than four times the F100S limit — and equipped with a standard 1-inch NPT rupture disk for overpressure protection.
The material selection directly affects both chemical compatibility and pressure/temperature envelopes. F100S and F100A share the 316L wetted path; F100H, F100B, and F100P all use C-22 nickel alloy.