Micro Motion F400 Flow and Density Meter
Micro Motion F400 Flow and Density Meter
Model:F400S999CQBZEZZZXCEQ
Manufacturer: Micro Motion
Size: 100*100*50cm
Detail Specification: Flow and Density Meter
Comment:Carton
Payment:T/T
HS CODE:9026201090
Delivery time: 8 weeks
product details
Micro Motion F400 Flow and Density Meter
The Micro Motion F400S999CQBZEZZZXCEQ is a 4-inch (DN100) Coriolis flow and density meter built on the F400S sensor platform — the largest size in Emerson’s compact, drainable F-Series line. The trailing “S” confirms a 316L stainless steel wetted path, delivering direct mass flow, volume flow, density, and temperature measurement from a single instrument.
The F400S is the highest-capacity F-Series sensor, engineered for high-flow-rate applications where a 4-inch line is standard. Its dual-tube, drainable architecture ensures complete product recovery during batch changeovers and CIP cycles. The meter’s ±0.20% mass flow accuracy and ±0.002 g/cm³ density accuracy are standard for this class, with optional upgrades pushing both specifications into custody-transfer territory. This configuration ships with a Micro Motion 2400S compact transmitter (the “EZZZXCEQ” suffix includes the transmitter and electronics package), creating a self-contained measurement package ready for field installation.
Technical Specifications — F400S999CQBZEZZZXCEQ
| Parameter | Specification |
| Base Sensor | F400S |
| Line Size | 4 in (DN100) |
| Wetted Material | 316L Stainless Steel |
| Flow Tube Configuration | Dual tube, drainable |
| Mass Flow Accuracy (Liquid) | ±0.20% of rate (standard); ±0.15% / ±0.10% optional |
| Mass Flow Repeatability (Liquid) | ±0.050% of rate |
| Volume Flow Accuracy (Liquid) | ±0.20% of rate (standard); ±0.15% / ±0.10% optional |
| Volume Flow Repeatability (Liquid) | ±0.050% of rate |
| Mass Flow Accuracy (Gas) | ±0.5% of rate |
| Mass Flow Repeatability (Gas) | ±0.25% of rate |
| Density Accuracy (Liquid) | ±0.002 g/cm³ (2.0 kg/m³) standard; ±0.001 g/cm³ optional; ±0.0005 g/cm³ optional |
| Temperature Accuracy | ±1 °C ±0.5% of reading |
| Sensor Maximum Working Pressure | 2,160 psig (149 barg) |
| Transmitter | Micro Motion 2400S Compact |
| Approvals | UL, CSA, IECEx, NEPSI, ATEX, Ingress Protection (IP66/67), EMC effects |
The ±0.20% liquid mass flow accuracy is the F400S baseline, but the meter is factory-configurable to ±0.10% for applications demanding tighter control. Gas flow accuracy at ±0.5% of rate with ±0.25% repeatability makes the F400S equally viable for natural gas and compressed gas measurement. The enhanced density calibration option at ±0.0005 g/cm³ (0.5 kg/m³) is available on the F400S, but not on high-temperature or high-pressure F-Series variants.
The F400S carries custody transfer approvals from MID OIML R 117, NTEP, Measurement Canada, and INMETRO Brazil — a credential set that opens doors to fiscal measurement and trade applications. Marine approvals include ABS, BV, CCS, DNV, LR, KR, NKK, and RINA. The sensor’s 2,160 psig (149 barg) pressure ceiling is significantly higher than the F100S and F150S models (both rated at 1,450 psig / 100 barg), reflecting the F400S’s heavier-wall construction for larger line sizes.
Model String Decoded: F400S999CQBZEZZZXCEQ
Every segment of the model number corresponds to a configuration decision:
F400 — F-Series compact, drainable platform; 4 in (DN100) line size
S — 316L stainless steel wetted parts. Per Emerson’s F-Series nomenclature, the trailing letter after the size code designates material or application: S for stainless steel, H for nickel alloy C22, A for high-temperature 316L, B for high-temperature nickel alloy, and P for high-pressure service
999 — Process connection configuration code for the DN100 flange and fitting arrangement
CQ — Process connection type and pressure rating code
B — Additional sensor configuration (orientation, drainability, or connection variant)
Z — Calibration and accuracy option code; standard calibration unless upgraded
E — Electronics or approval configuration code
ZZZ — Reserved placeholders for options not selected in this build
XCEQ — Transmitter and electronics package code; in this configuration, corresponds to the 2400S compact transmitter with the associated mounting and communication options
F-Series Material Options: S, A, H, B, and P Compared
The F400 platform is available in five wetted-material and application configurations:
F400S — 316L stainless steel, standard temperature and pressure. Maximum working pressure: 2,160 psig (149 barg). The most widely deployed variant for hydrocarbon liquids, water, chemicals, and general process fluids.
F400A — 316L stainless steel, high-temperature version. Extends the fluid temperature envelope for hot process streams.
F400H — Nickel alloy C22, standard temperature. C22 resists chlorides, oxidizing acids, and aggressive chemical media that would degrade 316L.
F400B — Nickel alloy C22, high-temperature version.
F400P — Nickel alloy C22, high-pressure version. Rated to 6,250 psig (430 barg) — more than four times the F400S limit — with a rupture disk for overpressure protection.
For the F400S999CQBZEZZZXCEQ configuration, the “S” confirms 316L construction at the 149 barg pressure ceiling. If the application involves aggressive chlorides, high-temperature caustics, or pressures exceeding 149 barg, an F400H or F400P alternative should be evaluated before specification.
Transmitter: Micro Motion 2400S Compact
The included 2400S transmitter is a compact, integrally mounted unit designed for continuous process control. Key capabilities include:
Display: Two-line LCD panel with standard user interface; glass or plastic lens options
Controls: Two local operation membrane buttons (accessible by removing the transmitter housing cover)
Output Variables: Mass and volume flow, product net content/flow, temperature, density, and concentration
Power: Self-switching AC/DC input with automatic voltage recognition; 85–265 VAC (50/60 Hz, 4 watts typical, 7 watts maximum) or 18–100 VDC (4 watts typical, 7 watts maximum); compliant with Low Voltage Directive 2006/95/EC and IEC 61010-1
Analog Output: Channel A — one 4–20 mA active output, non-intrinsically safe; Channel B — configurable as one active or passive frequency output, discrete output, or discrete input
Communication: HART/Bell 202 connection via two binding posts (requires removing transmitter housing cover); PROFIBUS-DP and DeviceNet options available (DeviceNet discontinued)
Mounting: Field-mount, integrally mounted to the F400S sensor
Application Types: Continuous process control, two-phase flow/entrained gas, meter verification
Hazardous Approvals: C1D2 Zone 2
The F400S sensor is also compatible with the 1600, 5700, and 1500/2500 transmitters for installations with different I/O or hazardous-area requirements. The 2400S supports Smart Meter Verification — an online diagnostic that confirms meter performance without interrupting the process.












