pressure transmitter,temperature Transmitter,flow meter-ST-YK CLD

YOKOGAWA K9802NA

YOKOGAWA K9802NA

Model:K9802NA

Manufacturer: YOKOGAWA

Size: 20*20*30cm

Detail Specification: Control CPU Board

Comment:Carton

Shipping Weight:0.8Kg

Estimated Lead Time: 12 Weeks

Payment:T/T

HS CODE:9026201090

Unit price (EXW): 3500US dollars

Delivery time: 20-24 weeks

product details

YOKOGAWA K9802NA

H1: Yokogawa K9802NA Main CPU Board for GC1000 Mark II Process Gas Chromatograph

Meta Description: Get full‑spec details for Yokogawa K9802NA main CPU control board, core spare component for GC1000 Mark II online gas chromatograph. Learn hardware specs, system integration, troubleshooting and field replacement for petrochemical analyzer maintenance.

The Yokogawa K9802NA is original OEM main CPU control hardware built exclusively for Yokogawa GC1000 Mark II process gas chromatograph platforms. Serving as the central computational and logic‑control backbone of the analyzer cabinet, the K9802NA manages detector signal acquisition, chromatogram peak processing, oven temperature sequence control, sample‑valve timing scheduling, local HMI interaction and multi‑protocol industrial data output. This plug‑in eurocard‑format circuit board is a mission‑critical spare part for end‑users operating legacy GC1000 Mark II units across refineries, natural‑gas treatment plants and chemical complexes. It restores full analytical functionality after main‑controller failure without cabinet‑level hardware retrofitting.

H2: Core System Functions of Yokogawa K9802NA

Yokogawa K9802NA undertakes end‑to‑end internal orchestration for the GC1000 Mark II analyzer. It digitizes raw analog signals transmitted from TCD or FID detector assemblies, executes baseline filtering, peak recognition, retention‑time locking and concentration computation based on stored analytical methods. The board generates precise timing triggers for pneumatic sample switching valves, governs PID oven temperature regulation loops, and continuously monitors interlock conditions such as carrier‑gas pressure, cabinet temperature and fault status. It mediates bidirectional data traffic across the GC proprietary backplane bus to I/O cards, communication modules and local touch‑screen terminal. Process measurement values, alarm events and chromatogram snapshots are buffered locally and pushed outward to plant DCS, PLC or remote maintenance workstations. All instrument configuration, calibration coefficients and audit logs persist on‑board to survive unexpected power cycling.

H2: Full Technical Specifications – Yokogawa K9802NA

H3: Computing & Memory Hardware

  • CPU Type: Industrial embedded processor optimized for chromatographic math operations and real‑time sequence control
  • On‑board Flash Memory: 128 MB non‑volatile flash, stores instrument firmware, analysis sequences, calibration tables
  • SDRAM Working Memory: 64 MB high‑temperature grade RAM for live chromatogram waveform buffering
  • FRAM Non‑volatile Storage: Integrated FRAM chip; preserves alarm history, user settings and calibration parameters through power loss
  • Firmware Compatibility: Matches official GC1000 Mark II firmware revisions; supports in‑situ firmware upgrade via USB or Ethernet
  • Method Storage: Supports multiple independent analytical method profiles for switching between different gas‑component measurement workflows

H3: Electrical & Physical Form Factor

  • Input Power: +24 VDC, fed from GC1000 Mark II internal cabinet power supply unit
  • Max Power Draw: 17.5 W under full analytical cycle loading
  • Mechanical Format: Standard plug‑in eurocard, designed for dedicated CPU slot on GC1000 Mark II system backplane
  • PCB Treatment: Conformal coated printed‑circuit board against dust, humidity and mild corrosive vapors inside analyzer houses
  • Net Weight: 1.9 kg
  • Installation Condition: For indoor cabinet mounting only; not for direct field hazardous‑area exposure

H3: Interface & Communication Parameters

  • Backplane Interface: Proprietary high‑speed parallel analyzer bus for inter‑board communication with AI, AO, relay and COM modules
  • Ethernet Port: 10/100 Mbps RJ45; Modbus‑TCP/IP, remote diagnostics, maintenance terminal access
  • Serial Port: RS‑485 physical layer supporting Modbus‑RTU for auxiliary peripheral connection
  • Local USB 2.0 Host Port: Used for firmware flashing, import / export of analysis methods, chromatogram and alarm‑log extraction
  • HMI Driver: Native signal output for driving GC1000 Mark II local operator touch‑screen panel
  • Digital Logic Handling: Processes discrete input signals for oven over‑temp interlock, carrier‑gas low‑pressure alarm, valve position feedback; generates control outputs for internal cabinet subsystems

H3: Environmental Operating Ratings

  • Operating Temperature Range (inside analyzer cabinet): 0 °C up to +55 °C
  • Relative Humidity: 5 % to 90 % RH, non‑condensing environment required
  • Vibration Resistance: 10‑150 Hz vibration tolerance compliant with typical process‑analyzer‑house mechanical conditions
  • Storage Temperature: ‑20 °C ~ +70 °C for spare‑part inventory storage
  • ESD Sensitivity: Requires anti‑static wristband and mat during handling and replacement

H3: Software‑Enabled Functional Specifications

  • Calculation Algorithms: Baseline correction, peak area integration, retention‑time shift compensation, external‑standard calibration, normalization calculation
  • Alarm Classification: Multi‑level fault reporting for communication drop‑out, signal drift, temperature deviation, gas‑supply anomaly; outputs coded diagnostic trouble codes
  • Data Logging: Circular buffer stores component readings, chromatogram traces and system event logs for audit review
  • User Permission: Tiered access rights for operator, maintenance and administrator‑level accounts
  • Diagnostic Capability: Board‑level self‑test routine on power‑on to detect memory, bus and internal circuit faults

H2: Standout Benefits of Genuine Yokogawa K9802NA

  1. Native hardware compatibility: The Yokogawa K9802NA follows original backplane bus timing and signal definitions; drop‑in replacement, no wiring or mechanical modification needed for GC1000 Mark II cabinets.
  2. Conformal‑coated PCB build improves service life under humid, dust‑prone analyzer‑house conditions common in refinery and petrochemical sites.
  3. Consolidated control architecture integrates chromatogram computation, sequence scheduling and communication functions on one card, reducing quantity of intermediate failure‑prone auxiliary boards.
  4. FRAM power‑fail safe storage minimizes re‑configuration work after unexpected power outages or board swap‑out.
  5. Convenient USB and Ethernet workflows simplify firmware updates, method migration and remote equipment troubleshooting for maintenance teams.
  6. Standard eurocard plug‑in design cuts instrument downtime during scheduled or emergency spare‑part change‑out under proper power‑off ESD‑controlled conditions.

H2: Real‑World Application Use‑Cases for Yokogawa K9802NA

  • Emergency spare‑part swap for failed main CPU controller on running GC1000 Mark II online gas chromatograph
  • Hardware refresh for aged GC1000 Mark II systems showing intermittent communication, boot‑up or signal‑processing faults
  • Post‑surge‑damage repair after cabinet power‑supply anomaly destroys original control card
  • Site spare‑part inventory for petrochemical and natural‑gas plants executing preventive maintenance strategies for analytical equipment
  • Rebuild and commissioning of pre‑owned GC1000 Mark II analyzer units for process‑gas monitoring projects

H2: Field Installation & Commissioning Best Practices

Shut down full GC1000 Mark II cabinet power before removing or inserting Yokogawa K9802NA. Apply strict ESD protection to prevent electrostatic damage to sensitive integrated circuits. Slide the board fully into the designated CPU backplane slot and fasten front‑panel locking screws securely. After power‑restore, upload matching firmware revision, import saved analysis methods and calibration datasets from USB storage. Execute cold system initialization, validate oven temperature control loops, sample‑valve sequence timing, detector signal response and all communication links to DCS / HMI infrastructure. Perform complete calibration using certified standard‑gas mixtures before returning the gas chromatograph to online production mode. Yokogawa‑trained service technicians are strongly recommended for on‑site board replacement and final validation work.

H2: Frequently Asked Questions

H3: Which analyzer model is Yokogawa K9802NA designed for?

Yokogawa K9802NA is purpose‑built exclusively for the Yokogawa GC1000 Mark II process gas chromatograph series. It is not directly compatible with other Yokogawa GC families including K9409GT or GC8000 without major hardware modification.

H3: Can existing analysis methods be reused after installing a new K9802NA?

Yes. Export all methods and calibration records to USB memory prior to board removal, then import these files to the fresh Yokogawa K9802NA unit. Even with imported data, full calibration with reference standard gas is still mandatory to satisfy metrology compliance requirements.

H3: What symptoms indicate potential K9802NA board failure?

Common failure indicators include analyzer boot hang‑up, touch‑screen unresponsiveness, missing chromatogram data, unstable DCS communication links, recurring system‑level fault codes that persist after verifying power supply, detector assemblies and peripheral I/O modules.

H3: Does Yokogawa K9802NA carry independent explosion‑proof certification?

No. The Yokogawa K9802NA is an internal plug‑in component. It must be housed within the original certified GC1000 Mark II analyzer cabinet for deployment inside hazardous classified plant zones. The board itself provides no intrinsic explosion protection.

Obtain relevant certificates and quotations for the product

Scroll to Top