pressure transmitter,temperature Transmitter,flow meter-ST-YK CLD

YOKOGAWA SC25V-AGP25-120

Model:ZR202G100SCETMEASCTF1

Manufacturer: YOKOGAWA

Size: 20*20*30cm

Detail Specification: Vortex Flowmeter

Comment:Carton

Shipping Weight:4.5Kg

Estimated Lead Time: 12 Weeks

Payment:T/T

HS CODE:9026201090

Unit price (EXW): 680 US dollars

Delivery time: 12 – 16 weeks

 

product details

YOKOGAWA SC25V-AGP25-120

The Yokogawa SC25V-AGP25-120 is a general-purpose combination pH sensor engineered for industrial process and wastewater treatment applications requiring reliable real-time pH monitoring. As Yokogawa’s first 12mm combination pH sensor featuring an external liquid earth electrode, this unit integrates a pH glass electrode, Ag/AgCl reference system, titanium liquid earth, and Pt1000 temperature sensor into a single compact housing-1. The sensor delivers consistent measurement stability across demanding environments where accurate pH control directly impacts product quality and regulatory compliance.

Measurement RangepH 0 to 14
Process Temperature (G-Glass)–10°C to 80°C (14°F to 176°F)
Process Temperature (L-Glass)+15°C to 130°C (59°F to 266°F)
Maximum Pressure0 to 10 bar (0 to 145 psi)
Min. Conductivity>10 µS/cm
Junction Resistance0.2 to 5 kΩ
Reference SystemDouble junction, Ag/AgCl with saturated KCl + Ag+ ion trap
Glass Impedance (Type G)200 MΩ at pH 7, 25°C
Glass Impedance (Type L)775 MΩ
Asymmetry Potential8 ±15 mV
Slope>96% (of theoretical value)
Sensor Shaft Diameter11.9 ±0.1 mm
Sensor Length120 ±2 mm
All Measuring ElementsWithin 25mm from sensor tip
Dimensions StandardDIN 19263:2007-05
Process ConnectionPG13.5 / M25 adapter
Wetted Materials: O-ringViton
Wetted Materials: Earth RingTitanium
Wetted Materials: JunctionPorous PTFE
Wetted Materials: Sensor BodyGlass / PEEK
Temperature SensorPt1000 to IEC 751
Response Time (pH step 7 to 4)<15 sec for 90%
Response Time (Temp step 10°C)<90 sec for 90%
Stabilization Time<120 seconds (0.02 pH unit/10s)
Storage Temperature–10°C to 50°C (14°F to 122°F)

Key Features

Large Electrolyte Reservoir – Increased KCl volume extends operational lifetime well beyond conventional 12mm pH sensor designs-17.
External Titanium Liquid Earth – Enhances measurement stability in low-conductivity and dirty process streams, reducing ground loop errors-1.
ATEX/IECEx Intrinsic Safety Certification – ATEX II 1G Ex ia IIC T3…T6 Ga. Permits safe installation in hazardous areas with gas groups IIA, IIB, and IIC-26. Ambient temperature compatibility for T6 (–40°C to +40°C), T4/T5 (–40°C to +55°C), and T3 (–40°C to +105°C)-26.
Wide Temperature & Pressure Capability – Handles processes from freezing to 130°C (266°F) at pressures up to 10 bar-1.
Integrated Pt1000 Temperature Compensation – Delivers highly accurate automatic compensation for pH measurement drift-17.
Porous PTFE Reference Junction – Large-surface-area diaphragm provides stable reference potential and resists fouling-24.
Low Conductivity Performance – Reliable measurement down to 10 µS/cm, suitable for high-purity water monitoring-17.
CIP & Steam Cleanable – Hygienic design supports in-place cleaning protocols without sensor removal-17.
Variopin Connector – Facilitates rapid sensor replacement and secure electrical connection-17.

Installation & Compatibility

For mounting in Yokogawa fittings, a PG13.5 to M25 adapter is available. The SC25V-AGP25-120 can be installed line, in a PR10 retractable fitting, FF20 flow fitting, FS20 subassembly, or FD20 immersion fitting-2. The sensor interfaces with intrinsically safe certified pH transmitters including the Yokogawa FLXA21 series, PH202S series, or Smart Adapter SA11-P1 (for ID-chip equipped versions)-26. Maximum cable length is 100 meters-.

Set/AGP Suffix Code System

The AGP25 suffix designates a general-purpose analog configuration for ATEX/IECEx environments, suitable for typical industrial processes, surface water monitoring, and general wastewater treatment-. For high-temperature, chemically aggressive applications, the SC25V-ALP25 variant is recommended. The Variopin-equipped combined sensor configuration (pH, reference, liquid earth, temperature) is fully contained within the 12mm shaft—all active measuring elements are located within 25mm of the sensor tip-26.

Electrical Data for Intrinsically Safe Circuits

ParameterValue (without ID-chip)Value (with ID-chip)
Voltage (Ui)18 V18 V
Current (Ii)170 mA170 mA
Power (Pi)0.4 W0.4 W
Inductance (Li)0 mH0 mH
Capacitance (Ci)0 nF0.4 nF

For intrinsically safe installations, the sensor must be connected to a certified associated pH transmitter that meets the specified input circuit restrictions. Connection to non-intrinsically safe equipment voids the hazardous area suitability-26.

Applications

Chemical processing (brine solutions, scrubber systems), power generation (flue gas desulfurization scrubbers), petrochemical separation (API separators in wastewater treatment, MEG/TEG processes), industrial effluent neutralization, surface water monitoring, food and beverage CIP circuits, high-purity water systems, and general wastewater treatment facilities.

 

Obtain relevant certificates and quotations for the product

Scroll to Top

Send Us a Message