product details

YOKOGAWA ZR402GMEEA

 

The YOKOGAWA ZR402GMEEA is a high-performance separate-type zirconia oxygen analyzer converter, engineered for precise monitoring and control of oxygen concentration in industrial combustion processes. Developed by Yokogawa Electric Corporation, this converter serves as the intelligent core of the EXAxt ZR series oxygen analysis system, working in conjunction with the ZR22G detector to deliver reliable, real-time oxygen measurements in demanding environments.

The YOKOGAWA ZR402GMEEA is specifically designed for combustion control applications across energy-intensive industries including steel manufacturing, electric power generation, petrochemical processing, ceramics production, pulp and paper, food processing, and textiles-53. By enabling complete combustion monitoring, the ZR402G series contributes significantly to energy conservation and environmental protection, helping reduce CO₂, SOx, and NOx emissions-13.

Model Code Breakdown: Understanding ZR402GMEEA

The model designation ZR402GMEEA follows Yokogawa’s systematic suffix code structure. Here is the detailed interpretation:

Code PositionDesignationMeaning
ZR402GBase ModelSeparate-type Zirconia Oxygen Analyzer Converter
MCable EntryM20 × 1.5 mm conduit connection
EDisplay LanguageEnglish display (LCD touchscreen interface)
EElectrical ConnectionStandard electrical termination
AOptional SpecificationStandard configuration without additional options

The YOKOGAWA ZR402GMEEA is a member of the broader ZR402G product family, which includes variants such as ZR402G-T-E-E-AZR402G-P-E-E-AZR402G-M-E-E-A/SCT, and ZR402G-M-E-E-A/C2/H-1. The “M” suffix specifies M20 cable entry threads, while the “-E” indicates English-language display firmware. The final “A” denotes the base specification without humidity analyzer configuration (/HS option) or special certifications.

Key Technical Specifications

General Specifications

The ZR402G separate-type converter features an intuitive LCD touchscreen display measuring 320 × 240 dots, enabling straightforward parameter configuration, calibration management, and oxygen concentration trend visualization-7-53. The display supports various functional screens including setting displays, calibration displays, and real-time trend graphs.

Input/Output Specifications

ParameterSpecification
Analog OutputTwo points, 4–20 mA DC, isolated
Maximum Load Resistance550 Ω
Contact OutputFour points (one fail-safe, normally open)
Contact InputTwo points
Auto-calibration OutputTwo points (for dedicated auto-calibration unit)
Digital CommunicationHART protocol (standard), 250–550 Ω

Power Requirements

ParameterSpecification
Rated Voltage100–240 V AC
Acceptable Range85–264 V AC
Rated Frequency50/60 Hz
Acceptable Frequency Range45–66 Hz
Power ConsumptionMax. 300 W; approx. 100 W for ordinary use

Environmental Specifications

ParameterSpecification
Ambient Temperature (Operation)–20 to +55 °C
Storage Temperature–30 to +70 °C
Ambient Humidity0–95% RH (non-condensing)
Installation Altitude2000 m or less
Installation Category (IEC 1010)Category II
Pollution Degree (IEC 1010)Degree 2
Enclosure ProtectionEquivalent to NEMA 4X / IP66 (outdoor installation)
Case MaterialAluminum alloy
Case ColorDoor: Moss green (Munsell 0.6GY3.1/2.0); Case: Off-white

Measurement Performance

The ZR402G series delivers exceptional measurement accuracy and responsiveness:

ParameterSpecification
Measurement ObjectOxygen concentration in combustion exhaust gas and mixed gas
Measurement SystemZirconium oxide (zirconia) sensor
Oxygen Concentration Range0.01–100 vol% O₂
Setting RangeAny range from 0–5 vol% O₂ to 0–100 vol% O₂
Display Range0–100 vol% O₂
Repeatability±0.5% of max. value (0–25 vol% O₂); ±1% of max. value (25–100 vol% O₂)
Linearity±1% of max. value (0–25 vol% O₂); ±3% (25–50 vol%); ±5% (50–100 vol%)
Drift±2% of max. value per month (zero and span)
Response Time90% response within 5 seconds
Warm-up TimeApproximately 20 minutes

Detector Compatibility and Probe Specifications

The YOKOGAWA ZR402GMEEA is designed to pair with the ZR22G separate-type zirconia oxygen detector. Key detector specifications include:

ParameterSpecification
Sample Gas Pressure–5 to +250 kPa (pressure compensation option required above +5 kPa)
Sample Gas Temperature (Probe only)0–700 °C
Sample Gas Temperature (with High-Temperature Probe Adapter)700–1871 °C
Probe Length Options0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
Probe MaterialSUS 316 (JIS)
Reference Gas Supply MethodsNatural air convection, instrument air, pressure compensated
Maximum Distance between Probe and Converter300 m (when using 1.25 mm² cable or equivalent)
Detector ConstructionHeater and thermocouple replaceable; non-explosion-proof
Detector ProtectionEquivalent to NEMA 4X / IP66
Probe Ambient Temperature–20 to +150 °C

Safety and Compliance

The ZR402G converter complies with major international safety and EMC standards:

Safety: Conforms to EN 61010-1:1993; CSA C22.2 No.1010-1 certified; UL 3111-1 certified

EMC: Conforms to EN 61326:1998

Wiring Connections: G1/2, Pg13.5, M20 × 1.5 mm, 1/2 NPT (eight conduit holes)

Installation Methods: Panel mounting, wall mounting, or 2-inch pipe mounting

 